Hologic

Panther System





Product Specification
Sample throughput 3.5 hours to first result.
Panther completes full testing, through to result, for up to 275 samples in 8 hours or 750 in 16 hours.
Sample Capacity 120 sample tubes can be loaded at once - in 8 racks with 15 samples per rack. Every 15 minutes a rack can be removed and new samples can be loaded, whenever you are ready.
Reagent Capacity 1 reagent kit per lane (up to 250 tests); 4 lanes available to load any combination of assays (up to 1000 tests). Kits can be reloaded as they are consumed without pausing testing. 2000-test onboard universal fluids capacity.
Waste Capacity 750 tests
Component Dimensions
(w x d x h)
Panther System:
122 cm (48 in) x 81.5 cm (32 in) x 175 cm (69 in)

UPS (Optional):
21.4 cm (8.4 in) x 41.0 cm (16.1 in) x 32.5 cm (12.8 in)
Weight Panther System:
363 kg (800 lbs)

UPS (Optional)
34.5 kg (76 lbs)
Sample Tube Barcode Types Code 39, Code 93, Code 128 (ISBT 128)
Interleaved 2 of 5, and Codabar
Environmental Requirements Ambient Temperature: 15 - 30°C
Relative Humidity: 20 - 85%
Electrical Requirements Electrical Input:
100-230 +/- 10% VAC, 50-60Hz, single phase

Current Input:
Minimum of 15 amp circuit (dedicated); 20 amp circuit (dedicated) if used with optional UPS.

Current Draw:
Instrument and ancillaries draw 400 watts in power-save mode to 1400 watts peak 100 VAC circuit draws 14 amps; 230 VAC circuit draws 6 amps.
System Output Heat Dissipation:
550 watts (1878 BTU/hour) during steady state.

Liquid Waste:
Waste fluids drain to removable containers

The highly evolved senses of a hunter

With true sample-to-result automation, adaptable workflow options and a growing assay menu, the PantherĀ® system is built for maximum efficiency. See what it can do for you in your lab for molecular diagnostic testing.



Flexibility Freedom Control to improve turnaround time

  • No batching restrictions
    Random, continuous access so you can load samples for any patient, in any order, at any time.
  • Run multiple assays simultaneously
    Run up to 4 assays on-board with the Panther system and up to 32 with the Panther FusionM system. Run up to 4 assays from a single patient sample.
  • Remove and reuse reagents as needed without pausing the Panther system
    Reduce reagent waste.


Freedom to improve turnaround time to increase productivity

  • Full, Sample-to-result automation
    Increases test throughput without increasing labor.
  • Walkaway freedom
    Less hands-on time than any comparable instrument.
  • Streamlined calibration
    Run controls and calibrators only once every 24 hours.


Control and reduce costs

  • Automated process checks
    Reduce invalid samples and/or runs.
  • Adaptable workflow
    Control workflow and costs with flexibility and automation.
  • Chain-of-custody
    Bi-Directional LIS connection for fully-automated processing and reporting.