Abbott

ALINITY CI-SERIES



The Alinity ci-series consists of compact, scalable systems to maximize throughput and efficiency, making today’s high-performing laboratories run at their best, today and into the future.


UNIFORMITY

Enable a truly harmonized healthcare system


FLEXIBILITY

Respond to the changing needs of the market and your community with the right capital investment


OPERATIONAL PRODUCTIVITY

Leverage lab technology and data to redefine and improve operational productivity


CONFIDENCE

Quality test results can save lives and reduce the financial impact to your hospital system


Product Specification
Alinity c Alinity i Alinity ci
System Dimensions (HxWxD) 134 x 119 x 117 cm/1.39 m2 134 x 119 x 117 cm/1.39 m2 134 x 199 x 117 cm/2.33 m2
Methods Photometric, Potentiometric Chemiluminescence Photometric, Potentiometric, Chemiluminescence
Maximum Throughput Up to 1350 TPH Up to 200 TPH Up to 1550 TPH
Throughput/m2 Up to 971 TPH/m2 Up to 144 TPH/m2 Up to 665 TPH/m2
Scalability Up to 4 modules controlled by one System Control Module (SCM)
Continuous Access of Reagents, Calibrators, Controls and Consumables Yes
Flexible Stat Options Prioritize single rack as needed or configure multiple fixed positions
Sample Types* Serum, plasma, urine, cerebrospinal fluid, hemolysate, whole blood Serum, plasma, whole blood, urine Serum, plasma, urine, cerebrospinal fluid, hemolysate, whole blood
Sample Capacity 150 150 300
Sample Bar Code Types Code 128, Standard Code 39, Interleaved 2 of 5, Codabar
Sample Result Storage 200,000
Dead Volume 50 μL (sample cup)
Sample Volume* 1.5–35 μL 2–200 μL Alinity c: 1.5–35 μL
Alinity i: 2–200 μL
Sample Probe Carryover ≤0.1 parts per million†
Reagent Capacity Up to 70 refrigerated reagent cartridges onboard plus patented ISE (Na+, K+, and Cl-) Up to 47 refrigerated reagent cartridges onboard Up to 117 refrigerated reagent cartridges onboard plus patented ISE (Na+, K+, and Cl-)
Reagent Type 100% liquid ready-to-use
Reagent Onboard Stability* 5–60 days 15–30 days For Alinity c: 5–60 days
For Alinity i: 15–30 days
Automated Onboard Calibrators and Controls* Yes Yes (controls only) Alinity c: Yes
Alinity i: Yes (controls only)
Calibration Frequency* 1–60 days 15–30 days For Alinity c: 1–60 days
For Alinity i: 15–30 days
Sample, Clot and Bubble Detection Yes
Reagent Pressure Monitoring Yes
Sample Interference Measurement Yes;
hemolysis, icterus, and lipemia
No Yes;
hemolysis, icterus, and lipemia (CC only)
On Board Maintenance Records Yes
Online Error Code Help Yes
Host Interface HL7 or ASTM
Remote Diagnostics AbbottLink
Weight 712 Kg 624 kg 1160 kg
Electrical Requirements SCM: 90–264 V, 16 amp
Each Instrument: 180–264 V, 16 amp
Water Requirements Average: 27 L/hr
Max‡: <30 L/hr
Average: <10 L/hr
Max‡: <30 L/hr
Average: ≤37 L/hr
Max‡: <60 L/hr
Heat Output (processing) Average 2005 Btu Average 1634 Btu Average 3639 Btu
Noise Level (1 m) Alinity c: 55.9 dBA
Alinity i: 63.4 dBA
Laboratory Automation Connection ACCELERATOR a3600
Download Brochure Alinity c Alinity i Alinity ci